supplied from an AC power source which uses a
separating transformer (in accordance with IEC
60989) or be provided with an additional protective
earth conductor.
•
If a portable multiple socket outlet assembly is used
as an AC power source, it must comply with IEC
60601-1-1. The assembly must not be placed on the
floor. Using more than one portable multiple socket
outlet assembly is not recommended. Using an
extension cord is not recommended.
An operator of the medical electrical system must not
touch non-medical electrical equipment and the patient
simultaneously. This may cause an unsafe electrical
shock to the patient.
Use of portable phones or other radio frequency (RF)
emitting equipment (that exceed electromagnetic
interference levels specified in IEC 60601-1-2) near the
system may cause unexpected or adverse operation.
Monitor operation when RF emitters are in the vicinity.
Use of other electrical equipment on or near this system
may cause interference. Verify normal operation of
equipment in the system before use on patients.
•
•
•
1 Introduction
2067226-001
1-19
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...