REV 5
DIRECTION 2173225-100
10-5
2. From clamp K" thecablecontinues to strap point M" and then to clamp point
H." See Illustration 10-4.
ILLUSTRATION 10-4
CLAMP POINT E
CLAMP POINT H
STRAP POINT M
STRAP POINT G
3. Thecathodecablecontinues from clamp point H" and strap point G," to clamp
point E" to the front panel in a figure-eight configuration, and then terminates
at thetransformer in thebaseof theunit.
D
The distances are referenced in Illustration 10-5.
ILLUSTRATION 10-5
CLAMP
POINT K"
17.00
CLAMP
POINT E"
STRAP
POINT G"
CLAMP
POINT H"
TUBE END
"0"
19.75
49.00
83.00
90.00
41.00
STRAP
POINT M"
STRAP
POINT J"
176.00
REF
Strap point J" is installed with collimator cable.
Summary of Contents for AMX 4+
Page 1: ...0 0 1 1 2 2...
Page 2: ......
Page 3: ...D D D D D D D D D D D D D D D D...
Page 4: ...D D D D D D D D...
Page 6: ...iv...
Page 8: ...vi...
Page 14: ...xii...
Page 18: ...xvi...
Page 32: ...1 14...
Page 48: ...3 14...
Page 84: ...5 10...
Page 106: ...7 12...
Page 112: ...8 6...
Page 116: ...9 4...
Page 131: ......
Page 132: ...3 2 2 1 0 3 5 0 2 0 4 0 2...