medTester 5000C
Operators Manual
2-4
4.
After entering date and time, insert a thin, non-metallic device in
the “clock set enable switch” slot (labeled on the right side of the
medTester) to depress the switch and enter the following menu
command to store this information:
STORE
F5
Press
Esc
to escape this procedure and to keep current settings.
Note
There is an INIT, F4, menu to initialize the date and time of a
new instrument. The “clock set enable switch” must be pressed
as in step 4 above, to save this initialization.
Audio Transducer
The medTester 5000C comes with an audio transducer that beeps whenever
data is entered from the keyboard or during an alert situation. You can adjust
the volume of the beeper:
1.
Enter menu commands:
UTIL
F5
BEEPER
F5
2.
To change the beeper volume of keyboard entries or to change
the volume of an alert, press either
KEYS
F1
or
ALERT
F2
3.
Press
VOLUME
F3
to increase the volume to the desired level for
keys or alert, depending upon which beeper key you pressed
above. Volume Levels are 0-6 with 0 inaudible and 6 the loudest.
4.
Test the beeper level by entering the menu command:
TEST
F4
5.
After entering beeper volume, enter the following menu
command to store this information:
STORE
F5
Press
Esc
to escape this procedure and to keep current settings.
Summary of Contents for medTester 5000C Operators
Page 6: ......
Page 32: ...medTester 5000C Operators Manual 1 20...
Page 40: ...medTester 5000C Operators Manual 2 8...
Page 46: ...medTester 5000C Operators Manual 3 6...
Page 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Page 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Page 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Page 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Page 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Page 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Page 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Page 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Page 70: ...medTester 5000C Operators Manual 4 24...
Page 108: ...medTester 5000C Operators Manual 6 16...
Page 114: ...medTester 5000C Operators Manual 7 6...
Page 124: ...medTester 5000C Operators Manual 8 10...
Page 178: ...medTester 5000C Operators Manual 11 14...
Page 196: ...medTester 5000C Operators Manual 12 18...
Page 210: ...medTester 5000C Operators Manual 13 14...
Page 256: ...medTester 5000C Operators Manual 15 28...