medTester 5000C
Operators Manual
9-8
Sternum Paddle.
The sternum paddle test is similar to the APEX paddle test.
Two (2) leakage measurements are taken, again with closed ground and with open
ground. Both measurements are taken under normal polarity and closed neutral
conditions. Before these tests execute, this message displays:
Press
READY
to begin the tests. Measurements are in µA RMS.
Energy Tests.
Energy tests execute in the low range and again in the high range.
Low Range
is defined as 0-50 joules, and
High Range
is greater than 50 joules.
Defibrillator autosequences are set at the factory so that low range energy step
tests execute before high range energy step tests. You can customize your
defibrillator testing so that this order is reversed. Prior to executing these tests,
this message appears in the medTester 5000C display:
When you are prepared to begin the tests, press
READY
. Next follow these
steps:
1.
Charge the EUT (the equipment under test—defibrillator
you're testing) to the energy level prescribed in the
medTester 5000C display.
2.
Discharge the defibrillator into the Impulse paddle contacts,
and press the down arrow key on the medTester. The reading
sent to the Impulse appears on the bottom medTester display
line with instructions for the next energy step displayed on the
top line. Here's an example display of this information:
Connect EXT red probe to STERNUM paddle
READY
F5
Select MEDT menu on Impulse
Set Impulse to LOW range
READY
F5
ENERGY STEP 2: Charge to
10 J: ready
STEP 1: ENERGY =
5.0 J
Summary of Contents for medTester 5000C Operators
Page 6: ......
Page 32: ...medTester 5000C Operators Manual 1 20...
Page 40: ...medTester 5000C Operators Manual 2 8...
Page 46: ...medTester 5000C Operators Manual 3 6...
Page 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Page 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Page 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Page 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Page 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Page 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Page 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Page 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Page 70: ...medTester 5000C Operators Manual 4 24...
Page 108: ...medTester 5000C Operators Manual 6 16...
Page 114: ...medTester 5000C Operators Manual 7 6...
Page 124: ...medTester 5000C Operators Manual 8 10...
Page 178: ...medTester 5000C Operators Manual 11 14...
Page 196: ...medTester 5000C Operators Manual 12 18...
Page 210: ...medTester 5000C Operators Manual 13 14...
Page 256: ...medTester 5000C Operators Manual 15 28...