medTester 5000C
Operators Manual
10-24
2.
SECONDARY RATE HIGH LIMIT: Enter the high limit for the
flow rate. 0000.0 turns off high limit checking.
3.
SECONDARY RATE LOW LIMIT: Enter the low limit for the
flow rate. 0000.0 turns off low limit checking.
4.
PRIMARY FLOW RATE: Enter the nominal flow rate, 0000.1 to
0999.0 ml/hr.
5.
PRIMARY RATE HIGH LIMIT: Enter the high limit for the
flow rate. 0000.0 turns off high limit checking.
6.
PRIMARY RATE LOW LIMIT: Enter the low limit for the flow
rate. 0000.0 turns off low limit checking.
7.
TEST END TYPE: Select
8.
MANUAL to end the test by pressing a key.
9.
TIMED to end the test after a specific time:
10.
END TIME: Enter the elapsed time that will end the
test, 00:01 to 99:59 (hours:minutes).
11.
VOLUME
to end the test after a specific volume:
12.
END VOLUME: Enter the volume that will end the
test, 0000.1 to 9999.0 ml.
For a PCA Test:
•
BOLUS VOLUME: Enter the bolus volume, 00.50 to 99.99 ml.
•
BOLUS VOLUME HIGH LIMIT: Enter the high limit for the
bolus volume. 00.00 turns off high limit checking.
•
BOLUS VOLUME LOW LIMIT: Enter the low limit for the
bolus volume. 00.00 turns off low limit checking.
•
LOCKOUT TIME: Enter the lockout time between boluses, 00:01
to 99:99 (minutes:seconds).
•
END TIME: Enter the elapsed time that will end the test, 00:01
to 99:99 (hours:minutes).
IPT-1
For the IPT-1, there are 2 types of tests available, steady flow and high
accuracy:
1.
FLOW RATE: Enter the nominal flow rate, 0000.1 to 0999.0
ml/hr.
2.
FLOW TEST TYPE: Enter STEADY for steady flow or HI ACC
for high accuracy.
3.
RATE HIGH LIMIT: Enter the high limit for the flow rate.
0000.0 turns off high limit checking.
Summary of Contents for medTester 5000C Operators
Page 6: ......
Page 32: ...medTester 5000C Operators Manual 1 20...
Page 40: ...medTester 5000C Operators Manual 2 8...
Page 46: ...medTester 5000C Operators Manual 3 6...
Page 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Page 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Page 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Page 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Page 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Page 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Page 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Page 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Page 70: ...medTester 5000C Operators Manual 4 24...
Page 108: ...medTester 5000C Operators Manual 6 16...
Page 114: ...medTester 5000C Operators Manual 7 6...
Page 124: ...medTester 5000C Operators Manual 8 10...
Page 178: ...medTester 5000C Operators Manual 11 14...
Page 196: ...medTester 5000C Operators Manual 12 18...
Page 210: ...medTester 5000C Operators Manual 13 14...
Page 256: ...medTester 5000C Operators Manual 15 28...