medTester 5000C
Operators Manual
11-10
Input Requested (I) Category
This checkitem tells the operator to check something or to do something and
then input a comment or data. After responding to the prompt, press
ENT
to
move to the next checkitem.
Both P&I (B) Category
This checkitem combines both pass or fail and input. The operator enters
either pass or fail (the result of that checkitem’s test) or comments that the
operator can input for that test.
A
FAIL
stores the character
#
in the checkitem record.
Remote Command (M) Category
There are two sub-categories of remote commands—remote commands
themselves and pause commands. You can insert remote commands as
checkitems in a checklist. In this example (
CLKE
) the medTester sets up a
measurement for case leakage current, external leads:
The message
EXTERNAL CASE LEAKAGE CURRENT
is a comment.
In addition, you can include a remote command that is sent through the
medTester 5000C to a tester or analyzer (like the Impulse 4000 Defibrillator
Analyzer) connected to the medTester's COM2 serial port. If the
medTester 5000C does not recognize the command as one of its own, it
considers it as an external command. The medTester passes these commands
out the COM2 port to the tester or analyzer. The medTester automatically
takes a reading from the analyzer, displays it, and stores it.
If an external remote command is followed by a number, the
medTester 5000C waits that number of seconds for a response. Otherwise it
waits five seconds.
LIGHTS & INDICATORS…CHECK OPERATION
DATA
F3
FAIL
F2
PASS
F1
CLKE
EXTERNAL CASE LEAKAGE CURRENT
COMMAND
Summary of Contents for medTester 5000C Operators
Page 6: ......
Page 32: ...medTester 5000C Operators Manual 1 20...
Page 40: ...medTester 5000C Operators Manual 2 8...
Page 46: ...medTester 5000C Operators Manual 3 6...
Page 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Page 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Page 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Page 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Page 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Page 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Page 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Page 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Page 70: ...medTester 5000C Operators Manual 4 24...
Page 108: ...medTester 5000C Operators Manual 6 16...
Page 114: ...medTester 5000C Operators Manual 7 6...
Page 124: ...medTester 5000C Operators Manual 8 10...
Page 178: ...medTester 5000C Operators Manual 11 14...
Page 196: ...medTester 5000C Operators Manual 12 18...
Page 210: ...medTester 5000C Operators Manual 13 14...
Page 256: ...medTester 5000C Operators Manual 15 28...