Biomedical Test System
medCheck Module
11
11-11
If an external command is followed by the number 0, the medTester will not
expect or look for a response from the external analyzer. It stores nothing in
the test record for that checkitem.
Note
Remote commands begin with alphabetic characters. See
Remote
Commands List
in Chapter 12,
Remote Operation
.
Pause Command (M) Sub-Category
You can also insert a checkitem in a checklist that simply creates a pause in
the execution of the checklist. A pause command is any numeric string (1 to
32768) that inserts a pause for that number of seconds. Here's an example:
The message
SECOND DELAY
is a comment. The seconds count down on the
display before the checklist continues.
medTester Autosequence (A) Sub-Category
Autosequences are placed in a checklist with the insertion of the
autosequence number. These numbers are listed by type—safety, defibrillator,
IV pump—in their respective chapters in this manual. When the
autosequence number appears in the checklist, the medTester 5000C executes
that autosequence. When executed as part of a checklist, these autosequences
do not prompt you for pretest information. Posttest prompts do appear. The
medTester stores checklist autosequence data separate from the checklist
record. Upon completion of the autosequence, the medTester returns to the
remaining checkitems in the checklist.
If You Make a Mistake
During the execution of the checklist you may notice that you have made a
mistake. For example, you could enter
pass
when you want to enter
fail
on the
result of a test checkitem. In this case, the medTester 5000C allows you to
back up in the checklist to a previous checkitem so that you can correct the
error. To back up in the checklist, press the up arrow key. You can then rerun
that checkitem.
With
M
(remote pause command) and
A
check items, you have two (2)
seconds to back up in the checklist before the checkitem executes.
10 SECOND DELAY
PAUSE:
10
SECONDS
REMAINING
Summary of Contents for medTester 5000C Operators
Page 6: ......
Page 32: ...medTester 5000C Operators Manual 1 20...
Page 40: ...medTester 5000C Operators Manual 2 8...
Page 46: ...medTester 5000C Operators Manual 3 6...
Page 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Page 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Page 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Page 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Page 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Page 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Page 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Page 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Page 70: ...medTester 5000C Operators Manual 4 24...
Page 108: ...medTester 5000C Operators Manual 6 16...
Page 114: ...medTester 5000C Operators Manual 7 6...
Page 124: ...medTester 5000C Operators Manual 8 10...
Page 178: ...medTester 5000C Operators Manual 11 14...
Page 196: ...medTester 5000C Operators Manual 12 18...
Page 210: ...medTester 5000C Operators Manual 13 14...
Page 256: ...medTester 5000C Operators Manual 15 28...