medTester 5000C
Operators Manual
13-10
Configure other autosequence ports as described here.
Configuring the Checklist Port
Whenever you execute a checklist in the medTester 5000C that contains
M
commands that are external to the medTester, they are directed from the
medTester to the Wedge checklist port. These external commands are
executed only by the instrument to which they pertain.
Prior to executing a checklist containing external (
M
) commands, always
ensure that you connect the instrument that can execute the command to the
checklist port. For example, a
30Z
command can only be executed in the
Fluke Biomedical 214B Patient Simulator, and the medTester 5000C will
direct that command to the checklist port. Not connecting the 214B to this
port will result in an error message appearing in the medTester 5000C display.
The default checklist port on the Wedge is
COM2-8
. Follow these instructions
to configure a different checklist port:
1.
From the medTester 5000C main menu,
MENU 1
, press:
UTIL
F5
2.
From the utility menu, press the medTester right arrow and enter
the command:
WPORTS
F5
3.
Press the medTester right arrow to access additional menu
options.
1.
Press:
CHECK
F3
to see this display:
4.
Press the medTester up arrow key to increment the number to the
port you want as the defibrillator autosequence port, or press the
down arrow key to decrement the number.
SET WEDGE PORT FOR CHECKLISTS: 8
STORE
F5
Summary of Contents for medTester 5000C Operators
Page 6: ......
Page 32: ...medTester 5000C Operators Manual 1 20...
Page 40: ...medTester 5000C Operators Manual 2 8...
Page 46: ...medTester 5000C Operators Manual 3 6...
Page 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Page 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Page 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Page 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Page 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Page 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Page 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Page 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Page 70: ...medTester 5000C Operators Manual 4 24...
Page 108: ...medTester 5000C Operators Manual 6 16...
Page 114: ...medTester 5000C Operators Manual 7 6...
Page 124: ...medTester 5000C Operators Manual 8 10...
Page 178: ...medTester 5000C Operators Manual 11 14...
Page 196: ...medTester 5000C Operators Manual 12 18...
Page 210: ...medTester 5000C Operators Manual 13 14...
Page 256: ...medTester 5000C Operators Manual 15 28...