medTester 5000C
Operators Manual
13-4
Wedge Port Configurations
Wedge Port
Name
(Default Wedge
Port No.)
Description
Non-Invasive Blood
Pressure (NIBP)
Autosequences
(7)
•
Execute medTester 5000C NIBP autosequences.
•
Specific to the CuffLink NIBP Analyzer.
•
Both preprogrammed and user-configured autosequence types.
•
See Chapter 17,
Non-Invasive Blood Pressure (NIBP) Module
.
Checklist
(8)
•
Used when executing equipment-specific checklist,
M
-type checkitems
and
any medTester 5000C
commands.
•
Especially useful for devices that do not have their own autosequences, e.g., the medSim 300B
Patient Simulator, the Model 214 B Patient Simulator, and the Model 217A Patient Simulator.
•
Please see the following for more information:
•
Chapter 11,
medCheck Module.
•
Chapter 12,
Remote Operation.
•
Operator’s manual or user’s guide section on serial port operation for the device under
medTester 5000C control.
Installing the Wedge
This section describes the procedure you use to install and connect the
Wedge to the medTester 5000C. It includes instructions for disassembling the
medTester 5000C’s feet and tilt bail, and attaching the Wedge to the
medTester 5000C and connecting required medTester 5000C-Wedge cables.
See Figure 13-1,
Wedge Installation
at the end of this section.
Installing the Wedge to the bottom of a medTester 5000C is a quick and easy
process. The only equipment needed to install the Wedge is a Phillips
screwdriver, #2-type.
Warning!
—UNPLUG THE INSTRUMENT POWER PLUG!
Service should only be performed by qualified personnel. There
are dangerous voltages present inside this instrument. Always
unplug the power cord before opening the case. Never plug in
the power cord while the case is open.
Summary of Contents for medTester 5000C Operators
Page 6: ......
Page 32: ...medTester 5000C Operators Manual 1 20...
Page 40: ...medTester 5000C Operators Manual 2 8...
Page 46: ...medTester 5000C Operators Manual 3 6...
Page 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Page 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Page 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Page 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Page 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Page 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Page 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Page 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Page 70: ...medTester 5000C Operators Manual 4 24...
Page 108: ...medTester 5000C Operators Manual 6 16...
Page 114: ...medTester 5000C Operators Manual 7 6...
Page 124: ...medTester 5000C Operators Manual 8 10...
Page 178: ...medTester 5000C Operators Manual 11 14...
Page 196: ...medTester 5000C Operators Manual 12 18...
Page 210: ...medTester 5000C Operators Manual 13 14...
Page 256: ...medTester 5000C Operators Manual 15 28...