Biomedical Test System
Installation
2
2-5
Enabling Modules
If it should become necessary to reinstall modules 2 through 14, perform the
steps described below. You need to install each module from its separate
diskette. Installation requires running a program on an IBM-compatible
personal computer.
Note
You
MUST
install the modules that you have received in
numerical order, starting with Module 2, or the lowest numbered
module you have. Installing modules in numerical order will lead
to a successful installation.
1.
Use the medTester 5000C COM1 port.
2.
Decide on a computer COM port—(COM1 through COM4).
3.
Decide on a baud rate—(300, 600, 1200, 2400, 4800, 9600, or
19200).
4.
To change the medTester baud rate from
MAIN MENU 1:
A.
Press
F5 UTIL
, then
F1 BAUD
;
B.
Select the COM1 port;
C.
Press
F4 BAUD
until the desired baud rate is displayed;
D.
Press
F5 STORE;
Press
Esc
twice to return to
MAIN MENU 1
.
Note
If Module 2 (RS-232/Printer) has not yet been installed, the
medTester baud rate is not selectable and can not be adjusted. It
is permanently set to 9600. After you install Module 2, you can
change the baud rate.
5.
Connect the computer COM port to the medTester COM1 port
(medTester to PC null modem serial cable).
6.
Run the computer in DOS mode (command prompt).
Summary of Contents for medTester 5000C Operators
Page 6: ......
Page 32: ...medTester 5000C Operators Manual 1 20...
Page 40: ...medTester 5000C Operators Manual 2 8...
Page 46: ...medTester 5000C Operators Manual 3 6...
Page 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Page 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Page 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Page 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Page 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Page 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Page 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Page 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Page 70: ...medTester 5000C Operators Manual 4 24...
Page 108: ...medTester 5000C Operators Manual 6 16...
Page 114: ...medTester 5000C Operators Manual 7 6...
Page 124: ...medTester 5000C Operators Manual 8 10...
Page 178: ...medTester 5000C Operators Manual 11 14...
Page 196: ...medTester 5000C Operators Manual 12 18...
Page 210: ...medTester 5000C Operators Manual 13 14...
Page 256: ...medTester 5000C Operators Manual 15 28...