Biomedical Test System
IV Pump Module
10
10-15
•
Wait for the IDA 4 Plus to display
PRIME
on channel 1. Using a priming
syringe, prime IV line into IDA 4 Plus until the IDA 4 Plus displays
•
After pressing
F5
(
START
), the medTester 5000C displays
START FLOW
FOR IDA 4P AUTO-START.
•
Next the IDA 4 Plus displays
AUTO
and waits for the flow to start. After
starting the flow on the IV Pump, the IDA 4 Plus should display
FLOW
followed shortly by flow data.
•
If the IDA 4 Plus displays
ERROR
during a test, it is likely that a bubble
was detected in the line, and the test is invalid. End the test, and repeat it.
Single Rate Test
•
For a single rate test, the medTester displays the elapsed time (ET=), the
remaining time (RT=) for a timed test, the remaining volume (RV=) for a
volume ended test, the average flow rate (A=), the instantaneous flow rate
(I=), and the cumulative volume (V=).
•
After 5 minutes, the medTester checks each average flow rate value with
the programmed limits. If the average flow rate is outside the limits, it is
displayed with a ‘#’.
•
You can wait for the test to end automatically, if it is a timed or volume
ended test, or you can end the test by pressing F5 END.
•
At the end of the test, the medTester again checks the average flow rate
with the programmed limits. If the average flow rate is outside the limits,
it is displayed with a ‘#’ with the opportunity offered to continue or
repeat the flow test.
•
The medTester displays the final flow rate test results. Press F5 NEXT.
Then you can repeat the test or continue on to the pressure test, if
selected.
Dual Rate Test
•
The dual rate test measures the delivery of two sequential fluid volumes
pumped at different flow rates by the IV pump. This fluid delivery
technique is typically referred to as “piggybacking” in the clinical
environment. NOTE: the initial higher fluid delivery (“Secondary”) must
be five times (5:1) the following lower (“Primary”) flow rate and must
minimally consist of 1.0 ml measured IV fluid
Summary of Contents for medTester 5000C Operators
Page 6: ......
Page 32: ...medTester 5000C Operators Manual 1 20...
Page 40: ...medTester 5000C Operators Manual 2 8...
Page 46: ...medTester 5000C Operators Manual 3 6...
Page 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Page 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Page 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Page 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Page 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Page 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Page 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Page 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Page 70: ...medTester 5000C Operators Manual 4 24...
Page 108: ...medTester 5000C Operators Manual 6 16...
Page 114: ...medTester 5000C Operators Manual 7 6...
Page 124: ...medTester 5000C Operators Manual 8 10...
Page 178: ...medTester 5000C Operators Manual 11 14...
Page 196: ...medTester 5000C Operators Manual 12 18...
Page 210: ...medTester 5000C Operators Manual 13 14...
Page 256: ...medTester 5000C Operators Manual 15 28...