Biomedical Test System
Remote Operation
12
12-7
Forcing A Port
An alternative to the above procedure is
forcing
a port. For example, if you
want COM1 to be the remote port, type or send
remote1
. This will force
COM1 to act as the remote port. To force COM2, type or send
remote2
.
If you have made a remote connection, the medTester responds by sending:
REMOTE MODE
out the remote serial port. At the same time, the
medTester 5000C display becomes blank.
Command Syntax Rules
•
Send commands in uppercase or lowercase letters.
•
Spaces are ignored.
•
Use the backspace character for editing.
•
Terminate commands with a carriage return (CR), or line feed (LF),
or both.
•
Commands with an equal sign (
=
) require that at least one parameter
be included with the command.
•
Use commas to separate multiple parameters.
Complete lists of medTester 5000C remote commands and parameters are
shown in the
Remote Commands List
section later in this chapter.
Responses to Commands
•
Every command you issue to the medTester returns a response to
the remote serial port.
•
Responses return character strings, numbers, or data files.
•
If no other response appears, an asterisk (
*
) appears to indicate that
the command has been executed.
•
Previous commands must execute and return their responses before
the medTester will accept further commands.
•
The null command (consisting only of terminating characters)
returns a question mark (
?
).
•
Command responses are in uppercase letters followed by a carriage
return and line feed.
•
Measurement reading responses appear as they do on the
medTester 5000C. Numbers may appear with or without a decimal
point and are followed by the unit of measurement abbreviation.
See the table that follows.
Summary of Contents for medTester 5000C Operators
Page 6: ......
Page 32: ...medTester 5000C Operators Manual 1 20...
Page 40: ...medTester 5000C Operators Manual 2 8...
Page 46: ...medTester 5000C Operators Manual 3 6...
Page 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Page 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Page 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Page 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Page 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Page 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Page 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Page 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Page 70: ...medTester 5000C Operators Manual 4 24...
Page 108: ...medTester 5000C Operators Manual 6 16...
Page 114: ...medTester 5000C Operators Manual 7 6...
Page 124: ...medTester 5000C Operators Manual 8 10...
Page 178: ...medTester 5000C Operators Manual 11 14...
Page 196: ...medTester 5000C Operators Manual 12 18...
Page 210: ...medTester 5000C Operators Manual 13 14...
Page 256: ...medTester 5000C Operators Manual 15 28...