Biomedical Test System
Autosequences
5
5-13
Test Tag Printer
After performing an autosequence or checklist (test procedure created in your
CMMS) with the “Print Test Tag” prompt enabled, the medTester will
prompt you to choose whether or not to print a test label. Attaching a Test
Tag label to a piece of equipment makes it easy to identify when that piece of
equipment was tested, who conducted the test and if desired, when the
equipment is due for its next test.
This section describes the setup and operation for printing Test Tags through
the medTester 5000C, using an optional Test Label Printer. Either the Brady
TLS PC Link (PN 2245515), the Fluke Biomedical Model TLP-1 or TLP-2
label printers are compatible with the medTester 5000C.
The Test Tag printer function will print up to five lines of information. Each
line describes a different parameter called a field and can hold up to 23
characters comprised of the field description, the output data itself and any
spaces created by the medTester.
Note
With the TLP-1 & 2, lines longer than the maximum number of
allowed characters will
wrap
to the next line on the label. For
example, if line number one is longer than 23 characters,
remaining characters will print on the second line. Since every
other line will print on the line below and if all five lines are
used, this will cause the fifth line to print on the next label of the
roll. With the TLS printer, each line longer than 23 characters
will simply be truncated
You can customize the printout by choosing any or all of the five fields listed
in the table below. For example, you could exclude the fourth line (
Due
). You
can not include other device or test information on the label.
Test Tag
Line Number
Field Description
Maximum Number of
Characters in the
medTester
Maximum Number of
Characters on the
Label Line
1:
HEADER: (or facility
name)
30 23
2:
CN: (equipment control
number)
16 16
3:
DATE: (current date)
8
8
(mm/dd/yy)
4:
DUE: (next test due date)
9
9
5:
OP CODE: (operator
code)
3 3
Summary of Contents for medTester 5000C Operators
Page 6: ......
Page 32: ...medTester 5000C Operators Manual 1 20...
Page 40: ...medTester 5000C Operators Manual 2 8...
Page 46: ...medTester 5000C Operators Manual 3 6...
Page 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Page 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Page 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Page 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Page 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Page 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Page 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Page 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Page 70: ...medTester 5000C Operators Manual 4 24...
Page 108: ...medTester 5000C Operators Manual 6 16...
Page 114: ...medTester 5000C Operators Manual 7 6...
Page 124: ...medTester 5000C Operators Manual 8 10...
Page 178: ...medTester 5000C Operators Manual 11 14...
Page 196: ...medTester 5000C Operators Manual 12 18...
Page 210: ...medTester 5000C Operators Manual 13 14...
Page 256: ...medTester 5000C Operators Manual 15 28...