medTester 5000C
Operators Manual
4-8
4.
Select the keyboard function key that corresponds to the lead you
wish to test.
5.
To begin the measurement, press
ISOV ON
F5
. The test begins,
and the
ISO VOLTS
annunciator lights. The reading is displayed for
up to 15 seconds. You can escape to terminate the test. The test
self-terminates after 15 seconds.
Note
If you receive abnormally high leakage current measurements
with V1-V6 leads tied together, you can inspect individual V
leads by removing the other leads. This procedure allows you to
isolate the lead with a high leakage current reading.
Equipment Current
See Figure 4-8,
Equipment Load Current
.
This test measures current used by the equipment under test. The EUT is
plugged into the
TEST RECEPTACLE
. The receptacle is fixed at normal polarity,
closed ground, and closed neutral. The range of this test is 20 amps RMS with
an accuracy of
±
5% of the range.
1.
Plug the EUT into the
TEST RECEPTACLE
.
2.
Enter menu commands:
MANUAL
F2
EQU CUR
F3
Resistance
These two resistance tests use the four-terminal test method. Resistance is
measured in a range of 2.000
Ω
with an accuracy of
±
1% of range with a test
current of 100 mA.
Summary of Contents for medTester 5000C Operators
Page 6: ......
Page 32: ...medTester 5000C Operators Manual 1 20...
Page 40: ...medTester 5000C Operators Manual 2 8...
Page 46: ...medTester 5000C Operators Manual 3 6...
Page 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Page 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Page 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Page 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Page 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Page 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Page 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Page 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Page 70: ...medTester 5000C Operators Manual 4 24...
Page 108: ...medTester 5000C Operators Manual 6 16...
Page 114: ...medTester 5000C Operators Manual 7 6...
Page 124: ...medTester 5000C Operators Manual 8 10...
Page 178: ...medTester 5000C Operators Manual 11 14...
Page 196: ...medTester 5000C Operators Manual 12 18...
Page 210: ...medTester 5000C Operators Manual 13 14...
Page 256: ...medTester 5000C Operators Manual 15 28...