Biomedical Test System
Remote Operation
12
12-11
Downloading
The medTester 5000C receives file download commands from the remote
computer in the form of a AUTOSEQDOWN or CHECKDOWN.
CHECKDOWN involves the transmission of three files. Each file follows
this sequence:
1.
medTester sends the acknowledgment response of a
*
to the
remote computer.
2.
The medTester receives the expected line count from the
remote computer.
3.
If the expected line count is a valid integer in the correct
range, the medTester next sends the expected line count back
to the remote computer. If the line count integer is invalid, it
sends the error message:
ERR=03,ILLEGAL PARAMETER
and
quits.
4.
The medTester receives the file from the remote computer.
5.
In CHECKDOWN the first line contains the file name. If the
medTester determines that the first line is incorrect, it sends
the error message:
ERR=23,WRONG FILE
and quits.
6.
medTester checks the line count against the expected line
count. If the count is correct, it sends the acknowledgment
response of a
*
. If the count is incorrect, it sends the error
message:
ERR=24,LINE COUNT WRONG
and quits.
7.
If the file is sent correctly, the remote computer sends the
medTester the acknowledgment response of
*
. If the remote
computer aborts the file transfer and knows the file was not
sent correctly, it sends the medTester the
ABORT
message.
Summary of Contents for medTester 5000C Operators
Page 6: ......
Page 32: ...medTester 5000C Operators Manual 1 20...
Page 40: ...medTester 5000C Operators Manual 2 8...
Page 46: ...medTester 5000C Operators Manual 3 6...
Page 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Page 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Page 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Page 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Page 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Page 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Page 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Page 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Page 70: ...medTester 5000C Operators Manual 4 24...
Page 108: ...medTester 5000C Operators Manual 6 16...
Page 114: ...medTester 5000C Operators Manual 7 6...
Page 124: ...medTester 5000C Operators Manual 8 10...
Page 178: ...medTester 5000C Operators Manual 11 14...
Page 196: ...medTester 5000C Operators Manual 12 18...
Page 210: ...medTester 5000C Operators Manual 13 14...
Page 256: ...medTester 5000C Operators Manual 15 28...