Running Pacer Autosequence Tests
To begin running pacer autosequence tests, connect your medTester 5000C
to your Impulse 4000. These two instruments are connected through the
COM2 or Wedge port on the medTester and the RS-232 serial port on the
Impulse 4000.
Note
Refer to Chapter 1,
General Information – Accessories
, or the current
Fluke Biomedical Price List for the correct serial cable part
number needed.
From the medTester 5000C Pacer menu, select the autosequence name for
the autosequence you wish to run. The name and number of the selected
autosequence appears on the display followed by the prompt:
If a safety autosequence has executed since power up of the medTester, you
can have the medTester use the previously input prompt information for the
current pacer autosequence, from this prompt:
USE PREVIOUS STARTING INFORMATION?
YES
F1
NO
F2
To accept previously input prompt information, press
YES
, and go to
Test
Sequence
, which follows this section. If you choose
NO
, go to the prompt
instructions in
Pretest Device Prompts
.
Press F5 for all device info prompts
Summary of Contents for medTester 5000C Operators
Page 6: ......
Page 32: ...medTester 5000C Operators Manual 1 20...
Page 40: ...medTester 5000C Operators Manual 2 8...
Page 46: ...medTester 5000C Operators Manual 3 6...
Page 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Page 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Page 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Page 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Page 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Page 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Page 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Page 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Page 70: ...medTester 5000C Operators Manual 4 24...
Page 108: ...medTester 5000C Operators Manual 6 16...
Page 114: ...medTester 5000C Operators Manual 7 6...
Page 124: ...medTester 5000C Operators Manual 8 10...
Page 178: ...medTester 5000C Operators Manual 11 14...
Page 196: ...medTester 5000C Operators Manual 12 18...
Page 210: ...medTester 5000C Operators Manual 13 14...
Page 256: ...medTester 5000C Operators Manual 15 28...