Biomedical Test System
Autosequences
5
5-15
1.
From the menu, press
F1
(
CONFIG
) as before to configure your
test tag (test tag lines on/off), or press
F2
(
PRINTER
) to select a
label printer.
2.
From the (
PRINTER
) menu press
F1
(
TLP
) to select Fluke
Biomedical TLP printers (TLP-1 or TLP-2), or
F2
(
BRADY
) to
select the Brady printer. The selection is displayed.
3.
Press
F5
(
STORE
) to store the selection in nonvolatile
EEPROM memory.
Note
When using the Brady printer, it is not necessary to set the
number of lines on the label with the
UTIL / PRTPAGE / TAGS
menu path. However, you still need to select the fields you want
printed on the label through the
CUSTOM
/
TAG
/
CONFIG
menu
path.
To configure a medTester serial port for test tag printing, see
Local Output for Records and Tags
in Chapter 12,
Remote
Operation
. To configure one of the Wedge adapter COM2 ports
for test tag printing, see
Setting Output Ports
in Chapter 13,
The
Wedge Adapter
Optional Monitoring Autosequences
Two optional autosequences are available in the Waveform/Extended Testing
Module. Both are designed as monitors.
•
Line monitor voltage test—periodically measures line voltage, and
•
Environmental test—which measures ground potential and
resistance between a common ground point and up to 99 other
ground points in a room.
MONITORING AUTOSEQUENCES
Autosequence
Name
Function
Menu Command
From MENU 2
Remote
Command
LINEMON Line
voltage
monitor.
LINEMON
F4
AUTO
F1
A16
ENVMON Ground
potential
and resistance.
ENVMON
F5
AUTO
F1
A17
Summary of Contents for medTester 5000C Operators
Page 6: ......
Page 32: ...medTester 5000C Operators Manual 1 20...
Page 40: ...medTester 5000C Operators Manual 2 8...
Page 46: ...medTester 5000C Operators Manual 3 6...
Page 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Page 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Page 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Page 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Page 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Page 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Page 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Page 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Page 70: ...medTester 5000C Operators Manual 4 24...
Page 108: ...medTester 5000C Operators Manual 6 16...
Page 114: ...medTester 5000C Operators Manual 7 6...
Page 124: ...medTester 5000C Operators Manual 8 10...
Page 178: ...medTester 5000C Operators Manual 11 14...
Page 196: ...medTester 5000C Operators Manual 12 18...
Page 210: ...medTester 5000C Operators Manual 13 14...
Page 256: ...medTester 5000C Operators Manual 15 28...