Biomedical Test System
ESU Module
14
14-7
8.
PHYSICAL INSPECTION—You have two physical inspection
lines, each of which can consist of a maximum of 40 characters.
After entering data for the first prompt, the second prompt
appears.
Note
See Chapter 5,
Autosequences
, for a complete explanation of
working with prompts. You can turn individual prompts on or
off and rename prompts six through eight (6-8) from the list
above. Chapter 6,
Customize Your medTester 5000C
, explains how
to customize prompts.
Test Sequence
After you input pretest prompt information, ESU tests execute as described
below:
1.
Set the ESU analyzer serial port baud rate to 2400.
The medTester uses the 2400 baud rate on COM2 for all
communications with the ESU analyzer during an ESU
autosequence.
2.
At the beginning of an ESU autosequence, the medTester
checks to make sure that an ESU analyzer is connected to
COM2. If none is detected, the user can still go through the
sequence in a simulation mode for demonstration purposes.
3.
Each autosequence can contain up to 30 steps. Each step
starts with a setup screen with the step number and type. Each
step can be one of 4 types. Each type executes as described in
the next section.
ESU Test Types
There are up to four types of tests that an ESU Analyzer can execute in
analyzing an ESU. The number of test types available are dependent on the
ESU Analyzer model and are described below.
Generator Output
This test measures the output of the ESU generator. It can use one of the
ESU analyzer internal loads or an external load module (the RF303
RS
can not
use external loads). The setup screen shows the load that will be used. If it is
an external load module, then that module must be installed on the ESU
analyzer.
Summary of Contents for medTester 5000C Operators
Page 6: ......
Page 32: ...medTester 5000C Operators Manual 1 20...
Page 40: ...medTester 5000C Operators Manual 2 8...
Page 46: ...medTester 5000C Operators Manual 3 6...
Page 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Page 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Page 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Page 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Page 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Page 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Page 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Page 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Page 70: ...medTester 5000C Operators Manual 4 24...
Page 108: ...medTester 5000C Operators Manual 6 16...
Page 114: ...medTester 5000C Operators Manual 7 6...
Page 124: ...medTester 5000C Operators Manual 8 10...
Page 178: ...medTester 5000C Operators Manual 11 14...
Page 196: ...medTester 5000C Operators Manual 12 18...
Page 210: ...medTester 5000C Operators Manual 13 14...
Page 256: ...medTester 5000C Operators Manual 15 28...