medTester 5000C
Operators Manual
1-14
Toolbox
Toolbox allows for the use of external measurement adapters for:
•
Tachometer—with a range of 100 to 20,000 RPM.
•
Temperature—with a range of 0 to 200
°
F or
0 to 100
°
C.
•
Humidity—ranging from 15% to 90% relative.
For more information about Toolbox, see Chapter 4,
Manual Tests
.
Test Receptacle
The medTester 5000C has a test receptacle on the top panel into which the
equipment under test (EUT) is plugged. The test receptacle supplies power to
the EUT. The EUT must be powered through the test receptacle whenever
you measure the following:
•
Leakage current—All case and ECG lead measurements.
•
Equipment current.
•
Power cord resistance.
Test receptacle power is normal except during case and ECG leakage tests
(not including isolation tests). For these tests you can set the power to the
following conditions, each of which is indicated by annunciator LEDs on the
top panel:
•
Ground—closed or open.
•
Polarity—normal or reversed.
•
Neutral—closed or open.
Note
During case internal leakage, ground is always open.
Current is supplied at a maximum instantaneous level of 20 amps.
CSA Label
On the back of the medTester 5000C, you can see a label which specifies the
conditions under which the test receptacle can operate:
•
1840 VA, 16 A continuous operation.
•
Up to 19 A for no longer than two (2) minutes. Then powered off
for eight (8) minutes (20 percent of duty cycle).
Summary of Contents for medTester 5000C Operators
Page 6: ......
Page 32: ...medTester 5000C Operators Manual 1 20...
Page 40: ...medTester 5000C Operators Manual 2 8...
Page 46: ...medTester 5000C Operators Manual 3 6...
Page 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Page 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Page 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Page 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Page 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Page 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Page 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Page 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Page 70: ...medTester 5000C Operators Manual 4 24...
Page 108: ...medTester 5000C Operators Manual 6 16...
Page 114: ...medTester 5000C Operators Manual 7 6...
Page 124: ...medTester 5000C Operators Manual 8 10...
Page 178: ...medTester 5000C Operators Manual 11 14...
Page 196: ...medTester 5000C Operators Manual 12 18...
Page 210: ...medTester 5000C Operators Manual 13 14...
Page 256: ...medTester 5000C Operators Manual 15 28...