PlasmaQuant MS Series
Plasma
Generation
75
Figure 6-18 - Below torch holder showing adjustment screw
6.
Screw alignment tool into the plasma coil so that the alignment pin doesn't
touch the coil.
Figure 6-19 - Proper positioning of alignment tool in plasma coil
7.
Position the alignment tool according to Figure 6-20 below.
8.
The upper left edge (left dashed line) must be parallel to the torch holder. The
gap must be horizontal on the upper side (green arrow).
9.
Alignment tool must sit flat on the torch holder as indicated by the red line in the
figure (i.e. alignment tool should not angle up or down).
10.
If a gap is visible (yellow arrow) plug in the tool from the left side into the coil.