Adjustment range for default alarm limits:
Fixed alarm limits / times:
Apnea pressure
after 15 seconds
Apnea flow
after 15 seconds
Apnea CO
2
after 15 seconds (after 60 seconds
in MAN/ SPONT)
– – : The alarm limit is deactivated as factory default.
NOTE: The newly set default alarm limits will be in
effect whenever the system is switched on or started
up from standby mode.
Certain alarm limits are switched off automatically in
MAN/SPONT: see tables, alarm limits for
MAN/SPONT, page 49.
Alarm limits
Adjustment range
Factory default
SpO
2
[%]
_
51 to 100
50 to 99
– –
92
Pulse
ª
[1/min]
_
31 to 300
30 to 299
120
50
EtCO
2
[mmHg]
_
1 to 75
0 to 74
50
– –
FiCO
2
[mmHg]
_
0 to 10
– –
5
– –
MV
[L/min]
_
0.1 to 39
0 to 38.9
12
3.0
FiO
2
[Vol.%]
_
19 to 100
18 to 99
– –
20
Fi Hal.
[Vol.%]
_
0.1 to 7
0 to 6.9
1.5
– –
Fi Iso.
[Vol.%]
_
0.1 to 7
0 to 6.9
2.3
– –
Fi Enf.
[Vol.%]
_
0.1 to 7
0 to 6.9
3.4
– –
Fi Des.
[Vol.%]
_
0.1 to 21.9
0 to 21.8
12.0
– –
Fi Sev.
[Vol.%]
_
0.1 to 9.9
0 to 9.8
3.4
– –
PAW
[cmH
2
O]
_
5 to 98
0 to 35
40
8
Configuring in Standby Mode
Setting Default Values
Operating Instructions Julian SW 3.n
97
Summary of Contents for Julian
Page 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Page 3: ...XXXXXXX ...
Page 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Page 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Page 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Page 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Page 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...