![Dräger Julian Operating Instructions Manual Download Page 143](http://html.mh-extra.com/html/drager/julian/julian_operating-instructions-manual_2526787143.webp)
C-Lock-ECG synchronization (optional)
Requirements for ECG
synchronization signal
Pos. pulse with voltage >4.5 V,
>10 ms duration for driving 2 mA
Max. permissible signal delay with
reference to current QRS complex
40 ms
Socket for 2-pin jack-plug,
dia. 3.5 mm
Jack layout
Signal isolation from other
electronic components
Dielectric strength
4 kV
Operating Data
Operating voltage
90 – 265 V, 45 - 65 Hz
Power input
400 W typically, max. 2.3 kW
Uninterruptible power supply
With fully charged battery: at least 30 minutes
(auxiliary outlets are not powered!)
Line power fuse
Automatic circuit-breaker 16 A
Auxiliary outlets
Each with two fuses 2 A DIN 41622 110/120 V AC
Medical gas supply
O
2
270 to 550 kPa (39 to 80 psi)
N
2
O
270 to 550 kPa (39 to 80 psi)
AIR
270 to 550 kPa (39 to 80 psi)
Max. cylinder pressure (O
2
and N
2
O)
19,000 kPa (2,750 psi)
Drive gas consumption
Operation
MV
±
1 L/min AIR or O
2
Standby and MAN/SPONT
0 L/min
Dimensions of Julian W x H x D
68 cm x 133 cm x 68 cm
Dimensions of the storage tray W x D
43 cm x 29 cm
Dimensions of the breathing system W x H x D 31 cm x 35 cm x 21 cm
Weight
Approx. 90 kg (200 lbs)
(ready for operation without
Vapor units and gas cylinders)
Weight of breathing system without soda lime
4.8 kg
Screen
Electro-luminescent screen, 10.4" diagonal
Protection class
Device
I, type B conforming to IEC 60601-1
SpO
2
sensor
Type BF
electrically isolated from protective earth
Electromagnetic compatibility EMC
Tested to IEC 60601-1-2
Technical Data
Interfaces
Operating Data
Operating Instructions Julian SW 3.n
143
m
Ground
Signal
Summary of Contents for Julian
Page 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Page 3: ...XXXXXXX ...
Page 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Page 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Page 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Page 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Page 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...