![Dräger Julian Operating Instructions Manual Download Page 80](http://html.mh-extra.com/html/drager/julian/julian_operating-instructions-manual_2526787080.webp)
Record
This function is used to determine the type of event
that triggers an entry in the numerical trend or an
online printout on the connected logging printer.
●
Select the line »record« with dial knob, press to
confirm.
The »record« menu appears on the left-hand side.
Display (example):
Time interval-started (min)
A list entry is activated after a fixed time interval in
minutes.
NiBP-started
An entry is made after every NiBP measurement with
new measured values. These values are read from an
external monitor which is connected to Julian via
MEDIBUS interface.
If no Dräger NiBP monitor is connected, no NiBP
entries will be activated.
warning-started
An entry is made whenever a Warning message is
triggered.
caution-started
An entry is made whenever a Caution message is
triggered.
●
Turn dial knob to select event triggering of entries,
press to confirm.
Activating monitor defaults
This function is used to activate the default monitoring
settings configured in standby mode which are active
whenever the system is switched on.
Alarm limits are not affected and remain at their
current settings.
●
Select the line »activate monitor defaults« with dial
knob, press to confirm, see page 92, "Setting Default
Values".
Monitoring
Configuring During Operation
Operating Instructions Julian SW 3.n
80
CMV
0 1 2 3 4 6 7 8 9
5
1 2 3 5 6 7 8 9
4
600 10
10
0
25
1.00
50
O %
2
L/min
PEEP
cmH O
2
Freq.
bpm
V
T
mL
T :T
IP
I
%
P
max
cmH O
2
Fresh Gas O + N O
2
2
PAW
40
20
0
settings
alarm sound
scale
parameter
record
activate monitor defaults
Configuration menu for List and Printer Entries
list entry
time
interval (min)
NiBP started
warn. started
caution started
printer entry
time
interval (min)
NiBP started
warn. started
caution started
1
2
5
10
off
on
off
on
off
on
off
1
2
5
10
off
on
off
on
off
on
off
calibration
alarms
pls. tone
adult
child
1:2.0
T
I
:T
E
MAC-config.
Summary of Contents for Julian
Page 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Page 3: ...XXXXXXX ...
Page 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Page 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Page 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Page 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Page 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...