Sensor Selection
Select sensor according to the following criteria:
– Patient weight
– Patient mobility
– Possible application site
– Patient perfusion
– Duration of use
The following table lists the specific sensor types
available and their characteristics. It is intended as an
aid for selection.
1) in unopened, undamaged packaging
Sensor type
OXISENSOR
OXISENSOR
OXISENSOR
OXISENSOR
OXISENSOR
I-20
D-20
DS-100 A
D-25
R-15
Age group
Infants Children
Adults
Adults
Adults
Patient
1 to 20 kg
10 to 50 kg
>40 kg
>30 kg
>50 kg
weight
Period of
Short and long-
Short and long-
Short-term
Short and long-
Short and long-
use
term monitoring
term monitoring
monitoring
term monitoring
term monitoring
Patient
Limited activity
Limited activity
Inactive
Limited activity
Inactive
mobility
patients only
patients only
Preferred
Toe
Finger
Finger
Finger
Nose
measuring
point
Sterility
1)
Sterile-packed
Sterile-packed
____
Sterile-packed
Sterile-packed
WARNING !
Use only original Nellcor SpO
2
sensors.
Follow Instructions for Use of the sensors –
incorrect positioning or use can cause tissue
damage.
Monitoring
SpO
2
Measurement
Operating Instructions Julian SW 3.n
86
Summary of Contents for Julian
Page 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Page 3: ...XXXXXXX ...
Page 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Page 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Page 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Page 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Page 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...