Connecting backup gas cylinders
for O
2
and N
2
O
Even if Julian is connected to a gas supply pipeline,
the cylinders must remain on the workstation as
backup supply.
●
On the back of the workstation:
place full tanks ( E-size) in the cylinder holders and
secure in position.
1 Attach pin index pressure reducing adapters onto the
cylinder valves.
2 Screw the gas hoses into the ports on the front of the
gas supply block.
3 Connect pressure sensor cable to the corresponding
sensor port.
●
Check cylinder pressures:
4 Slowly open the cylinder displays.
5 Check the pressure displays.
The pressure displays light up when the Julian
workstation is connected to line power.
WARNING !
●
Always use extreme caution when handling O
2
cylinders!
●
Oxygen intensely supports any burning!
No smoking, no open fire in areas where
oxygen is in use!
●
Always secure O
2
cylinders against tipping!
●
Do not expose O
2
cylinders to extreme heat!
●
Do not use oil or grease on O
2
equipment
such as tank valves or pressure regulators.
Do not touch with oily hands. Risk of explo-
sion!
●
Open and close valves slowly, with smooth
turns. Do not use any tools.
●
If a cylinder valve is leaky or difficult to
operate, it must be repaired by an authorized
professional.
●
Max cylinder pressure (O
2 and
N
2
O)
< 19.000 kPa (2750 psi).
Preparation
Connecting Supplies
Gas Connections
Operating Instructions Julian SW 3.n
34
N
2
O
1
1
1
O
2
2
2
3
4
4
Julian
D
5
Summary of Contents for Julian
Page 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Page 3: ...XXXXXXX ...
Page 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Page 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Page 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Page 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Page 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...