![Dräger Julian Operating Instructions Manual Download Page 107](http://html.mh-extra.com/html/drager/julian/julian_operating-instructions-manual_2526787107.webp)
●
Do not damage breathing hoses!
Dismantling the Suction System
1 Remove vacuum hose.
Refer to manufacturer's operating instruction for disa-
sembly and care of the suction equipment.
Dismantling the Breathing System
2 Fold up the writing top.
3 Pull catch and at the same time pull the slide-out
breathing system module out
CAUTION !
When removing and connecting breathing hoses,
always hold by the connector cuff and not by the
spiral reinforcement, otherwise the reinforcing
spiral may be torn off the cuff.
Breathing hoses with damaged spirals can easily
be kinked and may interrupt ventilation!
Always check the breathing hoses for damage
prior to use.
Damaged breathing hoses must not be used.
Care
Disassembly
Operating Instructions Julian SW 3.n
107
WARNING !
Always follow accepted hospital procedures
for handling equipment contaminated with
body fluids.