WARNING !
Disconnect power!
Before opening the unit, always disconnect
plug at the wall outlet first.
WARNING !
Preventive Maintenance work on the Julian
Anesthesia Workstation shall be performed by
trained and factory authorized staff only.
WARNING !
To avoid any risk of infection, clean and disin-
fect ventilator and accessories before any
maintenance according to established
hospital procedures - this applies also when
returning ventilators or parts for repair.
Maintenance
CAUTION !
In case of malfunction of this device, contact
your qualified local Dräger Service representa-
tive.
The device must be inspected and serviced (pre-
ventive maintenance) by competent and factory
authorized technical service representatives at
regular 6 month intervals. A record must be kept
on this preventive maintenance. We recommend
obtaining a service contract through your vendor.
Maintenance or repair of the Julian Anesthesia
Workstation shall be performed only by Dräger
qualified technical service representatives.
Care
Maintenance
Operating Instructions Julian SW 3.n
118
Maintenance Intervals
Preventive maintenance and service
By factory trained and authorized technical representatives every six
months
O
2
sensor
Replace when calibration is no longer possible or when the message
FIO2 INOP ! is displayed. Disposal, see page 120.
Flow sensor
Replace when calibration is no longer possible or when the message
FLOW INOP ! is displayed. May be incinerated at temperatures of
more than 800
°
C (1472 ºF).
Sample gas water trap
Replace when fill level reaches the mark or when the message
CO2 LINE BLK ! is displayed (assuming the sample line was not
otherwise blocked).
Cooling air filter, set of 3
Should be cleaned and thoroughly dried or replaced every month.
Must be replaced after one year at the latest.
Dispose of as domestic waste.
Lithium battery for data backup (2x)
Must be replaced by factory trained and authorized technical
representatives after two years.
Dispose of in accordance with all applicable waste disposal regulations.
Lead gel battery pack (2x)
Must be replaced by factory trained and authorized technical
representatives after two years.
Dispose of in accordance with all applicable waste disposal regulations.
Optical measuring system for
Must be checked by factory trained and authorized technical
determining the anesthetic
representatives every six months.
gas concentration
Time-keeper RAM
Must be replaced by professionals after four years.
Must be disposed of in accordance with the local waste disposal
regulations.
The supplier will make circuit diagrams, component part lists, descriptions and calibration instructions
available on request if needed for the maintenance of Julian.
Summary of Contents for Julian
Page 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Page 3: ...XXXXXXX ...
Page 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Page 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Page 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Page 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Page 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...