O
2
flush
1 Press »O
2
+« button.
2 The breathing bag inflates.
3 O
2
flow is audible.
Safety O
2
1 Turn Safety O
2
control valve counterclockwise.
2 The breathing bag fills up.
1 Set Safety O
2
back to » 0 «.
Preparing Julian for the self test
1 Close the Y-piece = plug firmly onto the cone.
2 Ensure that the sample line is not connected to the
patient circuit.
3 If necessary, seal the sample-line connector with
a luer-lock cap.
4 Set APL valve to MAN and adjust to 30 cmH
2
O.
●
Ensure that Safety O
2
flow control is set to » 0 «.
Preparation
Checks of Readiness for Operation
Operating Instructions Julian SW 3.n
41
2
1
3
4
3
2
1
2
1