The default settings for monitoring functions, as well
as fresh gas and ventilation parameters can be
configured in standby mode. Default settings for the
monitoring functions (pulse tone, scale, parameters,
record, interface, and traces) are in effect after the
system is switched on.
The default values for alarm limits, fresh gas
parameters and ventilation parameters are in effect
after each standby.
The basic configurations (language, time, date, and
tone sequence) remain until their next modification.
After switching the system on, the selection of
anesthetic agent is always set to "auto".
●
Switch Julian to standby mode.
●
Press »config.« menu key.
Display (example):
The following settings can be made in the menu:
– Default settings
– Calibration
– Anesth. agent
Fields highlighted by a gray background indicate
settings currently in effect.
The
z
z
z
z
symbol indicates a return to the previous
menu level.
●
Select, set and confirm new settings with dial knob.
Setting Default Values
Default values are in effect after the system is
switched on. Default monitoring settings can be
activated during operation with
»activate monitor defaults«,
see "Activating monitor defaults" on page 80.
●
Select and confirm default settings with dial knob.
Display (example):
The menu requests the user to enter a 4-digit
password in order to prevent unauthorized tampering
with the basic settings. This password is given to au-
thorized personnel during system training. This feature
can be disabled or the password be changed by
DrägerService on request.
Configuring in Standby Mode
Setting Default Values
Operating Instructions Julian SW 3.n
92
Standby / Configuration
calibration
defaults
O -sensor
2
21 vol.%
more
✓
alarm sound
scale
parameter
record
interface
alarm limits
curves
basic configurations
CMV settings
PCV settings
Fresh Gas Flow Control
0 1 2 3 4 6 7 8 9
5
1 2 3 5 6 7 8 9
4
gas
sensor
pls. tone
adult
child
✓
MAC-config.
MAC display
on
off
MAC correction
on
off
age 36
These default settings are
password protected and wil.
be active on startup.
To change the default settings,
enter the password:
0
1
2
3
4
5
6
7 8
9
Turn and confirm .
Standby / Configuration
* * *
calibration
O -sensor
2
21 vol.%
more
✓
defaults
gas
sensor
✓
MAC-config.
MAC display
on
off
MAC correction
on
off
age 36
Summary of Contents for Julian
Page 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Page 3: ...XXXXXXX ...
Page 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Page 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Page 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Page 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Page 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...