Anesthesia supported:
– Inhalation anesthesia in semi-closed to virtually closed
systems with "low flow" and "minimal flow"
techniques (for minimal gas and anesthetic agent
consumption) with a fresh gas flow of 0.5 to 12 L/min.
Ventilation Modes and Monitoring
Ventilation options:
– Automatic continuous mandatory ventilation (CMV)
and pressure-controlled ventilation (PCV)
– Manual ventilation (MAN)
– Spontaneous ventilation (SPONT)
The following measured values can be displayed:
– Peak pressure, mean pressure, plateau pressure and
PEEP
– Expiratory minute volume
Tidal volume V
T
Breathing rate
Patient compliance
– Inspiratory and expiratory concentration of O
2
, N
2
O,
anesthetic agent and CO
2
available options:
– Functional oxygen saturation (SpO
2
) and pulse rate
– Airway temperature
The following parameters can be displayed as
waveforms:
– Airway pressure
– Expiratory flow
– Inspiratory and expiratory concentration of O
2
, CO
2
and anesthetic agent
available option:
– Functional oxygen saturation (plethysmogram)
Graphical and numerical trends are also available.
Monitoring
is provided by adjustable alarm limits that can be
adapted to the ventilation mode.
Intended Medical Application
The Julian is indicated as a continuous flow anesthesia
system. The Julian may be used for manually assisted, or
automatic ventilation, and delivery of gases, anesthetic
vapor, and monitoring of oxygen concentration, breathing
pressure, and respiratory volume.
Federal law restrics this device to sale by or on the order
of a physician.
WARNING !
The Julian Anesthesia Workstation may only
be used under the supervision of qualified
medical personnel in order to provide immedi-
ate corrective action in case of a malfunction.
WARNING !
Do not use in conjunction with nuclear spin
tomography (MRT, NMR, or NMI)!
Equipment malfunction may result.
WARNING !
Do not use wireless or cellular phones within
33 feet (10 m ) of the equipment.
Radio communication devices may cause
malfunction in electromedical equipment.
WARNING !
DANGER, risk of explosion if used in the
presence of flammable anesthetics.
This device is neither approved nor certified
for use with combustible or explosive anes-
thetics (e.g. ether or cyclopropane).
WARNING !
To maintain grounding integrity, connect only
to a "hospital grade" receptacle. Always
disconnect supply before servicing.
WARNING !
Do not use accessories with the Julian Anes-
thesia Workstation that are not listed in the
component overview document enclosed with
the device.
Intended Use
Intended Medical Applications
Ventilation Modes and Monitoring
Operating Instructions Julian SW 3.n
18
Summary of Contents for Julian
Page 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Page 3: ...XXXXXXX ...
Page 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Page 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Page 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Page 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Page 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...