In the Event of Supply Failures
In the Event of a Power Failure
Julian will automatically switch to the integrated
uninterruptible power supply (UPS).
If the battery is charged, operation will be maintained
for at least 30 minutes.
The message POWER FAIL! appears on the screen
together with the remaining available battery capacity
in percent.
Example:
J
J
J
J
95 %
If the battery is almost completely discharged, the
message BATTERY LOW !!! appears on the screen.
Julian then automatically switches to MAN/SPONT
ventilation mode.
●
Adjust safety O2 flow control to a sufficient flow level
(1-12 L/min).
●
Check Vapor setting.
●
Ventilate patient manually!
In the Event of a Gas Failure
Julian generates an alarm if the O
2
, AIR or N
2
O
supply fails.
●
Open the valve on the corresponding backup gas
cylinder on the rear of the workstation if it had been
closed.
●
Check pressure display for the backup gas cylinder
●
Reestablish supply from pipeline.
If there is no backup supply of gas available, Julian
replaces the set gas mixture with AIR or O
2
:
Julian delivers 100 % AIR if the O
2
supply fails.
Julian delivers 100 % O
2
if the N
2
O supply fails.
Julian delivers 100 % O
2
if the AIR supply fails.
The amount of fresh gas flow (L/min) remains
constant.
WARNING !
Devices connected to the auxiliary outlets will
not be powered by the uninterruptible power
supply in the event of a power failure.
Operation
In the Event of Supply Failures
Operating Instructions Julian SW 3.n
61
N
2
O
O
2
CMV
SpO
2
et
CO
2
38
98
67
6.0
10
MV
Freq
600 10
10
0
25
1.00
50
alarm
limits
auto set
limits
alarm
info
list
config.
O %
2
L/min
PEEP
Freq.
V
T
mL
T :T
IP
I
%
P
max
Fresh Gas O + N O
2 2
curves
cmH
2
O
bpm
cmH
2
O
95%
POWER FAIL !
r
Pleth.
T
I
:T
E
1:2.0
O
2
45
41
2
N O
53
51
Fi
Fet
Iso.
0.8
0.6
Enf.
0.6
0.5
MAC (age: 36)
1.3
CO
2
40
0
PAW
0
20
Summary of Contents for Julian
Page 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Page 3: ...XXXXXXX ...
Page 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Page 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Page 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Page 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Page 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...