Fresh gas metering is still possible even in the event
of failure of one of the gases. For example, if the N
2
O
supply fails, AIR can be selected as the carrier gas.
Julian is equipped with two independent gas supply
failure alarm systems.
If the supply pressure of O
2
, AIR, or N
2
O drops
below 270 kPa (39 psi), the corresponding pressure
display on the front panel starts flashing. In the event
of an O2 supply failure, an uninterruptible continuous
audible alarm is generated.
If, in the event of inadequate gas supply (O
2
, AIR, or
N
2
O, fresh gas flow control is not able to deliver the
set fresh gas mixture, a corresponding warning is
displayed in the alarm field on the screen accompa-
nied by a continuous audible alarm.
An O
2
failure alarm is always signalled with the maxi-
mum alarm volume.
In the Event of a Major Hardware Fault
In this case, the message SYSTEM FAULT will be
displayed and an audible alarm is generated.
In the event of a hardware fault including the screen, the
audible alarm is generated only.
In these cases, Julian automatically switches to
MAN/SPONT ventilation mode.
No N
2
O will be delivered in this mode.
1 Adjust safety O
2
flow control to a sufficient flow level
(1-12 L/min).
2 Check Vapor setting.
●
Ventilate patient manually.
To silence the audible alarm:
●
Press dial knob.
In the event of system failure which also does not allow
an automatic switch to safe state the user may turn off
the Julian and ventilate the patient manually.
1 Adjust safety O2 flow control to a sufficient flow level
(1-12 L/min).
2 Check Vapor setting.
●
Ventilate patient manually!
Operation
In the Event of Supply Failures
Operating Instructions Julian SW 3.n
62
Julian
D
1
2
Summary of Contents for Julian
Page 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Page 3: ...XXXXXXX ...
Page 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Page 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Page 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Page 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Page 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...