![Dräger Julian Operating Instructions Manual Download Page 100](http://html.mh-extra.com/html/drager/julian/julian_operating-instructions-manual_2526787100.webp)
CMV Default Ventilation Settings
●
Select and confirm the line »CMV default settings«
with dial knob.
The menu keys for CMV are displayed:
Display (example):
To modify settings:
●
Press the respective menu key.
●
Set and confirm with dial knob.
To return to »CMV default settings«:
●
Press dial knob.
PCV Default Ventilation Settings
●
Select and confirm the line »PCV default settings«
with dial knob.
The menu keys for PCV are displayed:
Display (example):
To modify settings*:
●
Press the respective menu key.
●
Set and confirm with dial knob.
To return to the »PCV default settings« menu:
●
Press dial knob.
_____________
*
Maximum default setting for Pmax: 30 cmH
2
O.
Configuring in Standby Mode
Setting Default Values
Operating Instructions Julian SW 3.n
100
Standby / Configuration
Adjust default settings or confirm
600 12
10
0
25
PEEP
cmH O
2
Freq.
bpm
V
T
mL
T :T
IP
I
%
P
max
cmH O
2
calibration
O -sensor
2
21 vol.%
more
✓
defaults
alarm sound
scale
parameter
record
interface
alarm limits
curves
basic configurations
CMV default settings
PCV default settings
Fresh Gas Flow Control
0 1 2 3 4 6 7 8 9
5
1 2 3 5 6 7 8 9
4
gas
sensor
pls. tone
adult
child
✓
1:2.0
T
I
:T
E
MAC-config.
MAC display
on
off
MAC correction
on
off
age 36
Standby / Configuration
calibration
O -sensor
2
21 vol.%
more
✓
Adjust default settings or confirm ! Select default carrier gas using hard key.
defaults
alarm sound
scale
parameter
record
interface
alarm limits
curves
basic configurations
CMV default settings
PCV default settings
Fresh Gas Flow Control
0 1 2 3 4 6 7 8 9
5
1 2 3 5 6 7 8 9
4
gas
sensor
pls. tone
adult
child
✓
2.00
100
O %
2
L/min
Fresh Gas O
2
MAC-config.
MAC display
on
off
MAC correction
on
off
age 36
Summary of Contents for Julian
Page 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Page 3: ...XXXXXXX ...
Page 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Page 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Page 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Page 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Page 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...