Troubleshooting
Operating Instructions Julian SW 3.n
126
Message
Cause
Remedy
SPO
2
LOW
!!!
Oxygen saturation is below set lower
alarm limit.
Check ventilation.
Check O
2
concentration of fresh
gas.
Check SpO
2
sensor!
SPO
2
SEN DISC
!
SpO
2
sensor not plugged in.
Check sensor connection.
SPEAKER FAIL
!
* No alarm tone, loudspeaker faulty.
Call DrägerService.
SYSTEM FAULT
!!!
Internal device error.
WARNING Ventilate patient manually.
Do not operate the system.
Call DrägerService.
WATER TRAP
!
The collecting jar of the water trap at
the back of the unit is full.
Exchange the watertrap, WaterLock.
VENT ERR
!!!
The ventilator control unit is perfor-
ming an internal restart which takes
about 30 seconds. Julian is then res-
tored to its previous operating mode.
WARNING Ventilate patient manually.
Do not operate the system.
VOL ALRM OFF
!
* APNEA VOL or MIN VOL LOW
alarm disabled.
VOL ERR
!
* Flow sensor defective.
Replace flow sensor.
* Advisories related to hardware errors (e.g. sensor failures) are announced without audible signal.
Summary of Contents for Julian
Page 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Page 3: ...XXXXXXX ...
Page 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Page 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Page 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Page 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Page 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...