Select the desired position with the dial knob and
press to confirm.
Display (example):
The waveform is replaced and the cursor returns to
the
z
z
z
z
symbol in the selection menu.
To return to the standard page:
●
Press the dial knob,
or
●
press the
Q
key.
If a waveform field shall remain blank:
●
Select »blank« with the dial knob and press to
confirm.
Using the Volumeter Function
To observe and assess ventilation in manual and
pressure controlled ventilation modes and during
spontaneous breathing.
➀
Upper bar:
Current tidal volume (V
T
), preceded by its numerical
value.
The bar graph shows generation of V
T
➁
Lower bar:
Volumeter (minute volume measurement) with a mark
between each breath, preceded by its numerical
value.
The scale of the bar display depends on the default
setting, see "Configuring During Operation, Selecting
the scale" on page 78 and "Configuring in Standby
Mode, selecting the scale" on page 93.
Starting the volumeter
●
Press dial knob.
Values are deleted and volumeter is restarted if the
dial knob is pressed again before 60 seconds have
elapsed.
Time elapsed in seconds is displayed above the bar
with the integrated volume on the left. Individual
breaths are separated by segments in the bar. The
volumeter stops automatically with an acoustic signal
after 60 seconds.
Measured values are displayed for 4 minutes and then
disappear.
Setting
Tidal volume V
T
in L
Minute
volume MV
L/min
Scale for infants
and small children
0.2
2
Scale
for adults
1.0
10
Monitoring
Using the Volumeter Function
Operating Instructions Julian SW 3.n
76
MAN/SPONT
98
67
4.00
50
SpO
2
alarm
limits
CO -al.
2
off/on
alarm
info
list
curves
config.
Fresh Gas O + N O
2
2
O %
2
L/min
V
T
0.35
0
0.5
1
et
CO
2
38
6.0
10
MV
Freq
G
_
Volumeter running. To restart, confirm !
Volumeter
46 s
4.1
0
5
10
Pleth.
O
2
45
41
2
N O
53
51
Fi
Fet
Iso.
0.8
0.6
Enf.
0.6
0.5
MAC (age: 36)
1.3
CO
2
40
0
CMV
98
67
600 10
10
0
25
1.00
50
To select curve position, turn and confirm !
SpO
2
alarm
limits
auto set
limits
alarm
info
list
config.
O %
2
L/min
PEEP
cmH
2
O
Freq.
bpm
V
T
mL
T :T
IP
I
%
P
max
cmH
2
O
Fresh Gas O + N O
2 2
curves
et
CO
2
38
6.0
10
MV
Freq
Pleth.
T
I
:T
E
Flow
Volumeter
CO
2
1:2.0
O
2
45
41
2
N O
53
51
Fi
Fet
Iso.
0.8
0.6
Enf.
0.6
0.5
MAC (age: 36)
1.3
CO
2
PAW
Pressures
Pleth
O
2
AGas
Flow
FG/volu
blank
Volumeter
FG/press
➀
➁
V
T
0.35
0
0.5
1
Volumeter
46 s
4.1
0
5
10
Summary of Contents for Julian
Page 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Page 3: ...XXXXXXX ...
Page 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Page 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Page 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Page 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Page 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...