Flow measurement
(hot wire anemometry)
Tidal volume VT
Range
0.02 to 9.99 L
Resolution
0.01 L
Accuracy (under calibration conditions
Better than
±
8 % of the measured value or 0.01 L, the larger value
and 1013 hPa)
applies
Minute volume MV
Range
0 to 99.9 L/min
Resolution
0.1 L/min
Accuracy Better
than
±
8 % of the measured value
(under calibration conditions and BTPS*)
Respiration rate
Range
0 to 60 per minute
Resolution
±
1 per minute
Accuracy
±
1 per minute or
±
10 %, the smaller value applies.
Breathing gas temperature measurement
(NTC resistance)
Measuring range
20 to 50
°
C (68 to 122
°
F)
Resolution
1 K
Accuracy
±
0.5 K in the range from 30 to 41
°
C (86 to 106
°
F)
CO
2
measurement
(infrared spectrometry)
CO
2
measurement
(infrared spectrometry)
Sample rate
60 mL/min or
(selectable)
200 mL/min
Delay time for sampling
Less than 1 s (for a sample rate of 200 mL/min)
Response time t
10...90
at 200 mL/min
300 ms
at 60 mL/min
1 s
Measuring range
0 to 9.9 kPa (0 to 80 mmHg)
Resolution
0.1 kPa
Accuracy
±
0.2 vol% or 5 % of measured value
___________
* BTPS: Body temperature, current atmospheric pressure, gas saturated
Technical Data
Measuring Functions
Operating Instructions Julian SW 3.n
140
Summary of Contents for Julian
Page 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Page 3: ...XXXXXXX ...
Page 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Page 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Page 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Page 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Page 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...