Setting Fresh Gas Defaults
●
Select and confirm the line
»Fresh Gas Flow Control« with dial knob.
The menu keys for fresh gas settings are displayed:
Display (example):
To modify settings:
1 Press keys »N
2
O« or »AIR« to select default carrier
gas, press dial knob to confirm.
2 Press »O
2
%« menu key.
Set default O
2
concentration with dial knob,
press to confirm.
3 Press »L/min« menu key.
Set and confirm default fresh gas flow with dial knob.
To return to the »Fresh Gas Flow Control« menu:
●
Press dial knob.
Configuring in Standby Mode
Setting Default Values
Operating Instructions Julian SW 3.n
101
Standby / Configuration
calibration
O -sensor
2
21 vol.%
more
✓
Adjust default settings or confirm ! Select default carrier gas using hard key.
defaults
alarm sound
scale
parameter
record
interface
alarm limits
curves
basic configurations
CMV settings
PCV settings
Fresh Gas Flow Control
0 1 2 3 4 6 7 8 9
5
1 2 3 5 6 7 8 9
4
gas
sensor
pls. tone
adult
child
✓
2.00
100
O %
2
L/min
Fresh Gas O
2
MAC-config.
MAC display
on
off
MAC correction
on
off
age 36
Julian
1
2 3
Summary of Contents for Julian
Page 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Page 3: ...XXXXXXX ...
Page 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Page 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Page 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Page 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Page 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...