Cleaning /Replacing Cooling Air Filters
–
Clean filters monthly
1 Remove all three cooling air filters from their holders.
●
Clean in warm water to which a detergent has been
added; dry thoroughly.
●
Insert cooling air filters into their holders without
creasing.
●
Replace cooling air filters after not more than one
year. Dispose of filters as domestic waste.
Replacing Water Trap
– When fill level reaches the mark
or
– when the message
CO
2
LINE BLK ! is displayed
(if the sample line is correctly installed and free of any
blockage).
●
Hold water trap by its ribbed surface and pull it out of
its holder.
●
Push a new water trap into the holder until it clicks
into place
●
Dispose of old water trap as domestic waste.
Replacing O
2
Sensor
– When the message
FIO
2
INOP! is displayed
or
– when the sensor can no longer be calibrated.
1 Unscrew sensor holder containing O
2
sensor.
2 Remove the spent O
2
sensor from the holder and
insert a new O
2
sensor.
1 Screw holder back in its place.
●
For disposal of spent O
2
sensors, see page 120.
Care
Maintenance
Operating Instructions Julian SW 3.n
119
1
2
1
1
1
Summary of Contents for Julian
Page 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Page 3: ...XXXXXXX ...
Page 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Page 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Page 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Page 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Page 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...