MAC configuration
Anesthetic agent identification and display
Julian automatically identifies the anesthetic agent used
and adjusts the measurement and monitoring of the
anesthetic gas concentration to suit the gas identified.
If there is a mixture of two volatile anesthetic agents,
the concentration of the secondary anesthetic agent is
displayed from a MAC value of 0.1 MAC upwards. The
gas with the higher expiratory MAC value is displayed
above the secondary gas.
A secondary anesthetic agent becomes the main
anesthetic agent if its MAC value exceeds the MAC
value of the main anesthetic agent by 0.2 MAC.
A mixture of more than 2 volatile anesthetic agents
cannot be reliably detected.
Calculating the MAC values
Julian calculates the MAC value (Minimum Alveolar
Concentration) from the linear total of volatile anesthetic
agents measured and N
2
O. The measured values used
are the end-expiratory concentrations. Partial pressures
are taken into account.
Age-corrected MAC values are calculated according to
W. W. Mapleson, British Journal of anesthesia 1996, pp.
179-185.
Displaying MAC
To activate the MAC display:
●
Select column »MAC config.« by turning the rotary
knob and press to confirm.
●
Select »MAC display« by turning the rotary knob
and press to confirm.
●
Select »on« by turning the rotary knob and press
to confirm.
To select the age correction for the MAC display:
●
Select »MAC correction« by turning the rotary
knob and press to confirm.
●
Select »on« by turning rotary knob and press to
confirm.
Monitoring
MAC configuration
Operating Instructions Julian SW 3.n
83
SpO
2
CMV
SpO
2
et
CO
2
38
98
67
6.0
10
MV
Freq
600 10
10
0
25
1.00
50
alarm
limits
auto set
limits
alarm
info
list
config.
O %
2
L/min
PEEP
Freq.
V
T
mL
T :T
IP
I
%
P
max
Fresh Gas O + N O
2 2
curves
cmH
2
O
bpm
cmH
2
O
1:2.0
O
2
45
41
2
N O
53
51
Fi
Fet
Iso.
0.8
0.6
Enf.
0.6
0.5
MAC (age: 36)
1.3
T
I
:T
E
CO
2
40
0
PAW
0
20
600 10
10
0
25
1.00
50
O %
2
L/min
PEEP
cmH O
2
Freq.
bpm
V
T
mL
T :T
IP
I
%
P
max
cmH O
2
Fresh Gas O + N O
2
2
PAW
40
20
0
Confirm to select Isoflurane.
O
2
-Sensor
21 Vol.%
gas sensor
CMV
calibration
pulse tone
alarm sound
scale
parameters
record
activate monitor defaults
0 1 2 3 4
5
6 7 8 9
1 2 3
4
5 6 7 8 9
adult.
child
alarms
settings
✓
✓
T
I
:T
E
1:2.0
default
CO
2
on
off
SpO
2
on
off
HLM on
off
FiCO
2
5
MAC display
on
off
MAC correction
on
off
age 36
MAC-config.
Summary of Contents for Julian
Page 1: ...Julian Anesthesia Workstation Operating Instructions Software 3 n XXXX ...
Page 3: ...XXXXXXX ...
Page 6: ...Operating Instructions Julian SW 3 n 6 This page intentionally left blank ...
Page 16: ...Operating Instructions Julian SW 3 n 16 This page intentionally left blank ...
Page 64: ...Operating Instructions Julian SW 3 n 64 This page intentionally left blank ...
Page 136: ...Operating Instructions Julian SW 3 n 136 This page intentionally left blank ...
Page 144: ...Technical Data APL Valve Characteristics 144 APL Valve Characteristics ...