3.9 Clock and Lock Mode (-LKM/-LKS Option)
Clock and lock mode operation of two FCS18 AC power sources is available only if the –LKM
and –LKS options have been installed at the factory. With these options installed, it is possib
to lock an auxiliary
le
unit (-LKS) to a master unit (-LKM). The master unit controls the frequency.
six
e inputs. Do not connect these BNC’s between
two master units (-LKM’s) or damage to the unit could result.
Note: The maximum number of auxiliary units (-LKS) per master (-LKM) is one. No T
This configuration can be used to create multiphase power systems such as split-phase or
phases. The auxiliary unit must be set to external clock mode from the Control screen. See
section 4.2.5.
Two BNC connectors are provided on the rear panel of the FCS for clock and lock mode. Both
need to be connected between the master and auxiliary unit. On the master unit (-LKM), both
are outputs. On the auxiliary unit (-LKS), both ar
connectors are allowed to increase the number of auxiliary units as this may
cause reflections on the clock and lock signals.
Error! Reference source not found.
for the required connections between the –LKM
and –LKS units. The example is shown for two units, one master, one auxiliary.
WARNING:
DO NOT CONNECT THE AC OUTPUTS OF THE –LKM AND –LKS UNITS
TOGETHER. CLOCK AND LOCK OUTPUTS CANNOT BE PARALLELED TO
OBTAIN HIGHER OUTPUT CURRENTS.
Do not use clock and lock mode to obtain higher power capability on the same phase(s). For
higher power configurations, use the multi-chassis configuration through the system interface
connection instead. Refer to section 3.5.3 for multi-chassis configuration and connection
information.
The frequency of the auxiliary unit will track that of the master. The output phase angle of phase
A will be locked to the auxiliary unit as well to within 3°. This allows split phase or six phase
configurations to be created.
3.9.1 Configuration settings
Units configured with the –LKM option will show the Clock as INT (internal) and the mode as
CLK/LOC on the CONTROL screen. Units configured with the –LKS option can be set to INT
(internal) or EXT (external) clock from the CONTROL screen. The MODE setting on the
CONTROL screen of the –LKS unit determines the power on state for the clock setting. When
set to STAN (Stand-alone operation), the unit powers up with INT clock. When set to CLK/LOC
mode, it powers up in EXT clock mode suitable to clock and lock system operation. See section
4.2.5 for details.
3.9.2 Frequency measurements on –LKS units
FCS models configured with the –LKS option used in a clock and lock configuration will not
accurately measure frequency if the programmed frequency of the master unit (-LKM) exceeds
2000 Hz unless the frequency setting of the –LKS is set to the a value close to the programmed
frequency of the –LKM master unit. Setting the –LKS programmed frequency when it is
operating as a clock and lock auxiliary does not affect its actual output frequency as it is
controlled by the
–LKM master unit. This provides the controller with the required information to accurately
measure the frequency.
California Instruments
45
Summary of Contents for FCS Series II
Page 4: ...User Manual Rev B FCS Series II California Instruments 4...
Page 27: ...User Manual Rev B FCS Series II California Instruments 27...
Page 89: ...FCS Series II 89 Figure 5 1 FCS System Block Diagram User Manual Rev B California Instruments...
Page 95: ...User Manual Rev B FCS Series II Figure 6 1 Calibration Setup California Instruments 95...