User Manual – Rev B
FCS Series II
The EUT measurement data collected
separate tabs that can be displayed by
for the 101 EUT performance test is divided among four
clicking on each tab at the top of the test window.
Available tabs are:
Tab
Description
Parameters
This tab displays parametric measurement data for all phases (three-
phase mode) or phase A only (single-phase mode). Data displayed
includes Frequency (Hz), Voltage RMS (V), Current RMS (A), Apparent
he MIL-
ill be displayed on the
right. The ITHD limit is set in the “MS704_Distortion_Limits.ini” text file
Power (VA), Inrush current (A), Power Factor and Total Current
Distortion (%). The total VA for all three phases combined is shown for
three phase EUT’s. Any unbalance between phases is calculated and
displayed next to the total VA readout. If the unbalance exceeds t
STD 704 limit of 3%, this field will have a red background.
The current distortion is compared to the user specified ITHD limit value
shown to the right of the current distortion readout. If one of more of
phase currents exceeds this limit, a Fail result w
the
located in the LxGui application directory. (See paragraph 11.8.3.1)
Waveforms
This tab displays at least one full cycle of both the voltage and current
waveforms. (all waveforms are for L-N voltage and phase current). In
hase
is displayed at a time but all three phases are always acquired at the
e data for all three phases as well.
The Settings button can be used to change display scaling and graph
three-phase mode, the phase to be displayed can be selected by using
the phase selection radio buttons to the left of the graph. Only one p
same time. The test report will includ
colors if needed.
Curr Graph
This tab displays the current harmonic spectrum in graphical form (Ba
Chart - absolute). In three-phase mode, the phase to be displayed can
be selected by using the phase selection radio buttons to the left of the
graph. Only one phase is displayed at a time but all three phases are
always acquired at the same time. The test report will include data for a
three phases as well.
r
ll
The Settings button can be used to change display scaling and graph
colors if needed.
Curr Spectrum
This tab displays the current harmonic spectrum in tabular form (Numer
– absolute and relative to fundamental). In three-phase mode, the ph
to be di
ic
ase
splayed can be selected by using the phase selection radio
buttons to the left of the graph. Only one phase is displayed at a time but
phases are always acquired at the same time. The test report
will include data for all three phases as well.
f
t
This data is also part of the test report.
all three
The user can define relative limits for one or more current harmonics. I
set, the LxGui will compare actual harmonic current components agains
the user defined limits and display pass or fail in the right hand column.
The current harmonic limits can be set by editing the
“MS704_Distortion_Limits.ini” text file located in the LxGui application
directory. (See paragraph 11.8.3.1)
The EUT performance measurements are taken at the start of a test run and again after the
eeded by using the “File Read” button in the lower left corner of
the Parameters Tab. The last measurement data taken by the LxGui program is preserved in
selected test or tests have been completed. If a variable frequency test group is selected, the
data will be taken at 360 Hz, 400 Hz, 600 Hz and 800 Hz. Data for each frequency is saved and
can be recalled for display if n
California Instruments
194
Summary of Contents for FCS Series II
Page 4: ...User Manual Rev B FCS Series II California Instruments 4...
Page 27: ...User Manual Rev B FCS Series II California Instruments 27...
Page 89: ...FCS Series II 89 Figure 5 1 FCS System Block Diagram User Manual Rev B California Instruments...
Page 95: ...User Manual Rev B FCS Series II Figure 6 1 Calibration Setup California Instruments 95...