User Manual – Rev B
FCS Series II
4.2.9.1 CALIBRATION Menus
routine calibration procedures and equipment
requirements. Do not attempt calibration without consulting the user manual.
This menu also contains the LCD viewing angle adjustment.
The measurement calibration menu can be used to perform routine calibration of the internal
measurement system. The recommended calibration interval is 12 months. To enter the
calibration screens, the calibration password must be entered first.
Note: Refer to chapter 6 for details on
ENTRY
DESCRIPTION
VIEW ANGLE
0
-10 to +1
LCD viewing angle adjustment.
CAL PWORD
V range
Calibration password required to access all
is the high voltage range value. [270] The
password can be entered using the
Voltage
calibration screens. The calibration password
shuttle followed by the
ENTER
key.
Measurement Calibration Screens
MVOLT F/S
± 9999
Calibration coefficient for full-scale voltage
measurement.
MCURR F/S
± 9999
Calibration coefficient for full-scale current
measurement.
Output Calibration Screens
VOLT F/S
± 9999
Calibration coefficient for full-scale voltage
output. (calibrated by measurement)
CURR F/S
± 9999
Calibration coefficient for full-scale current
ouptut. (calibrated by measurement)
California Instruments
71
Summary of Contents for FCS Series II
Page 4: ...User Manual Rev B FCS Series II California Instruments 4...
Page 27: ...User Manual Rev B FCS Series II California Instruments 27...
Page 89: ...FCS Series II 89 Figure 5 1 FCS System Block Diagram User Manual Rev B California Instruments...
Page 95: ...User Manual Rev B FCS Series II Figure 6 1 Calibration Setup California Instruments 95...