User Manual – Rev B
FCS Series II
SXF 106 Tests
Voltage distortion spectrum test require the use of additional external equipment. In particular, a
high bandwidth (100Khz) AC source, a high frequency coupling transformer and coup
networ as sho
tandard. For this test, the AC s
e
can be programmed to a steady state condition of 400 Hz and Vnominal of 115 Vrms or
Vrms.
ling
k
wn in figure SXF106-1 of the Mil-Std 704 s
o c
230
ur
d SXF-106 Test Setup.
SXF 107 Tests – R
Figure 11-12: Require
evision F
SXF107
Total Voltage Distortion Rev F
Condition
Duration (s)
A
0.05 % VTHD
1800
F = 100%
H3 = 2.75%
H5 = 2.75%
H7 = 1.97%
H9 = 1.53%
H11 = 1.25%
H13 = 1.06%
H15 = 0.92%
SXF 108 Tests
SXF108 are DC offset tests and are not supported by the FCS Series AC Power sources.