User Manual – Rev B
FCS Series II
FREQUENCY TRANSIENTS
(A(CF) only)
Test 1
Seq. No.
Volt/Frequency
Time
1 115/400
5
minutes
2 115/350 5
sec.
3 115/320
0.2
sec.
4 0/320
0.2
sec.
5 115V/400
10
sec.
Test 2
Seq. No.
Volt/Frequency
Time
1 115/400
5
minutes
2 115/480
0.2
sec.
3 115/440 5
sec.
4
0/440
0.2 sec.
5 115V/400
10
sec.
FREQUENCY TRANSIENTS
(Group 1 only)
Seq. No.
Frequency
Time
1 400
5
minutes
2 480 5
sec.
3 400
10
sec.
4 320 5
sec.
5 400
10
sec.
Table 8-14: Abnormal Frequency Transient
e at 115V and will remain at this voltage through out the test except
he test will cycle the frequency three times as shown in Table 8-14.
1 and test2 for the A(CF) category are done in
This test will set the voltag
for the A(CF) category. T
Each repeat will start from sequence 2. Test
succession as a single test.
California Instruments
127