8.6 EMERGENCY TEST
From the DO160 MENU scroll to the EMERGENCY AC entry using the up and down cursor
keys. Press the ENTER key to select the EMERGENCY screens. The screen will appear as
shown in Figure 8-8.
F
8-8:
mergency
reens
The E
SCREEN
e following
1
VOLT FREQ MIN
igure
E
Sc
MERGENCY
has th
tests:
2
VOLT FREQ MAX
3 VOLT
UNBALANCE
The above tests can be selected by scrolling to the highlighted selection using the up and down
key and the ENTER key to start the selected test.
VOLT FREQ MIN
Standard/Group
RTCA
A(CF)
A(NF)
A(WF)
1
Ф
100 100 100 100
Voltage
3
Ф
101.5 101.5 101.5 101.5
Frequency
360 360 360 360
Standard/Group
Group1
Group2
Group3
1
Ф
104 104 104
Voltage
3
Ф
105.5 105.5 105.5
Frequency
360 360 360
Table 8-8: Emergency Voltage and Frequency Minimum
Standard/Group
RTCA
Group1
Group2
Group3
1
Ф
122 122 122 122
Voltage
3
Ф
120.5 120.5 120.5 120.5
Frequency
440 440 650 800
Standard/Group
Group1
Group2
Group3
1
Ф
122 122 122
Voltage
3
Ф
120.5 120.5 120.5
Frequency
440 650 800
Table 8-9: Emergency Voltage and Frequency Maximum
This test will set the voltage and frequency to a level defined by Table 8-8. The test will last for
30 minutes. The test will be repeated using the voltage from Table 8-9 and frequency from
Table 8-8. The
←
key (backspace) will terminate the test at any time.
California Instruments
123
Summary of Contents for FCS Series II
Page 4: ...User Manual Rev B FCS Series II California Instruments 4...
Page 27: ...User Manual Rev B FCS Series II California Instruments 27...
Page 89: ...FCS Series II 89 Figure 5 1 FCS System Block Diagram User Manual Rev B California Instruments...
Page 95: ...User Manual Rev B FCS Series II Figure 6 1 Calibration Setup California Instruments 95...