Brooks Automation
146
278580 Revision A
8 Service and Troubleshooting
LF80C ZeniD RFID Reader
8.4 Error Codes
Product Manual
8.4.3
Stream Function
disturbing field at
transmitting
frequency
check antenna surroundings for
possible disturbing sources
(monitors, servo motors, ...)
antenna is damaged
or too close to metal
exchange antenna, verify
antenna installation
SSAC
K
Name
Description
Possible cause
Correcting action
Stream
Functio
n
Description
Possible cause
Correcting action
S9F1
unrecognized DeviceID
message with wrong Devi-
ceID was sent to the
reader
send a message with the correct
DeviceID (can be taken from the
S9F1 message)
S9F3
unrecognized stream type
message with a unknown
stream type was sent to
the reader
check stream function syntax of
the message
S9F5
unrecognized function type
message with a unknown
function type was sent to
the reader
check stream function syntax of
the message
S9F7
illegal data
wrong RIC at reset mes-
sage
check RIC value
S9F9
Transaction Timer Timeout
host system does not
acknowledge the mes-
sage or sends no answer