Brooks Automation
84
278580 Revision A
7 Operation
LF80C ZeniD RFID Reader
7.3 Operation of SECS Protocol
Product Manual
Description
Used to identify the data requested.
The DATASEG corresponds to the page number (PAGEID) of multipage,
read/only and read/write transponders.
"00"
First page of any type of transponder or first page of
DATA
area in case of a multipage transponder.
Multipage-transponder (page 1 to page 17):
Start the reading or writing on the following page of a multipage
transponder:
...
"01"
page 1
"81"
Locked page 1
... ...
"11"
page 17 "91"
Locked page 17
Read/only transponder
"F0"
Read only the one page
Read/write transponder
"F1"
Read or write only the one page
Where used
S18F5, S18F7
Acknowledge code for new reader constant
0 ...
Parameter was set successfully
1 ...
Parameter could not be set
Where used
S2F16
Parameter number of reader (see data item ECV)
Where used
S2F13, S2F15
Reader parameter definition.
The values are displayed as decimal values, see
Î
Parameters.
DATASEG
Format:A[2]
EAC
Format: B[1]
ECID
Format: U1
ECV
Format: U1