Brooks Automation
278580 Revision A
95
LF80C ZeniD RFID Reader
7 Operation
Product Manual
7.3 Operation of SECS Protocol
Parameter 7 (0x07)
TARGETID high byte
High byte of the predefined TARGETID.
The TARGETID is changeable.
The default value is delivered from the serial number.
The following example shows the relation between serial number and
default TARGETID: serial number „0203MIS04660“
The last 5 characters of the serial number are to be interpreted as a
decimal device counter and they are used for the default TARGETID. In the
example, the decimal number „04660“ can be interpreted as a hexadecimal
value 0x1234. The default value for the TARGETID has the value „1234“
and the parameter 7 has the default value 0x12.
Parameter 8 (0x08)
TARGETID low byte
Low byte of the predefined TARGETID.
The TARGETID is changeable.
The default value is delivered from the serial number.
The following example shows the relation between serial number and
default TARGETID: serial number „0203MIS04660“
The last 5 characters of the serial number are to be interpreted as a
decimal device counter and they are used for the default TARGETID. In the
example, the decimal number „04660“ can be interpreted as a hexadecimal
value 0x1234. The default value for the TARGETID has the value „1234“
and the parameter 8 has the default value 0x34.
Parameter 9 (0x09)
Heartbeat time
The reader offers the option of generating a regular heartbeat. This means
the reader sends an S1F1 message to the host at the defined interval.
0 … No heartbeat
1 … 255 10 s (10 s - 2,550 s)
Default: 0 no heartbeat