CHAPTER 13: Talk and Test Modules for the SCAN-LAN VI
I
NDEX OF
R
EFRACTION
Set the Index of Refraction (IOR) each time you test a cable from a different
manufacturer or spool.
Attach the launch cables to TX (transmit) and RX (receive) ports on each handset
as shown in
Section 13.5
. Note that TX on one unit connects to RX on the other.
• Set the Main Unit dial to CAL.
• Highlight
IOR
with the
↑↓
keys.
• Press CALIBRATE (F4).
You can use system-default IOR settings, manually input new settings, or calculate
the correct IOR based on a known length of cable.
• Press DEFAULT (F4) to use the default settings, or
• Press DEC (F2) or INC (F3) to manually change the IOR setting, or
• To set IOR based on a known length of cable: Connect a cable of known
length to TX and RX on the main unit.
S
ETTING
L
OSS
B
UDGET
CALIBRATE IOR
DEFAULT
F1
F2
F3
F4
Press Enter To Save
INC
DEC
START
568A Horizontal Lnk
NVP:
1.490
CALIBRATE
IOR
SCAN-LAN IV
CALIBRATE
F1
F2
F3
F4