116
SCAN-LAN VI
6. Press SAVE (F4) or ENTER to save the new value.
7. To repeat, press Start (F1).
8. Press ESC to exit the NVP calibration mode.
10.1.2 U
SING
M
ANUFACTURER
’
S
C
ABLE
S
PECIFICATIONS
If you do not have a known length of cable available, but can identify the
cable manufacturer and part number, you may be able to determine the
rated cable NVP and load that NVP value into the SCAN-LAN VI. Many
common cable NVPs are listed in
Appendix B, Common NVP Values
.
To load a known NVP value:
1. Set dial to CAL.
2. Select NVP from the calibration menu and press CALIBRATE (F4)
or ENTER.
3. Press INC (F3) or DEC (F2) to adjust the NVP value.
4. Press ENTER to save the new NVP value. A message indicates
that the new value has been entered.
F1
F2
F3
F4
F1
F2
F3
F4